Device | Mode4-150 | Circumstance | AAMI4-151 | BHS4-152 | |||
Accutracker II | Aus | At rest | P/P | A/C | |||
CH-DRUCK (103) | Aus | At rest | P/P | A/A | |||
Daypress 500 | Osc | At rest | P/P | A/B | |||
DIASYS 200 | Aus | At rest | P/P | C/C | |||
DIASYS Integra | Aus | At rest | P/P | B/A | |||
Osc | At rest | P/P | B/B | ||||
ES-H531 | Aus | At rest | P/P | A/A | |||
Osc | At rest | P/P | B/B | ||||
Medilog ABP | Aus | At rest | P/P | ||||
Meditech ABPM-04 | Osc | At rest | P/P | B/B | |||
Nissei DS-240 | Osc | At rest | P/P | B/A | |||
OSCILL-IT | Osc | At rest | P/P | C/B | |||
Pressurometer IV | Aus | At rest | F/F | C/D | |||
Profilomat | Aus | At rest | P/P | B/A | |||
In pregancy | P/P | B/C | |||||
Profilomat II | Osc | At rest | F/F | C/B | |||
QuietTrak | Aus | At rest | P/P | B/B | |||
P/P | A/A | ||||||
Different posture | A/A | ||||||
During exercise | A/A | ||||||
In children | A/A | ||||||
In the elderly | A/A | ||||||
In pregnancy | F/F | B/B | |||||
A/A | |||||||
In pre-eclampsia | F/F | D/D | |||||
Save 33 Model 2 | Osc | At rest | P/P | B/B | |||
Schiller BR-102 | Aus | Passed | P/P | B/B | |||
Osc | Failed | P/P | B/B | ||||
SpaceLabs 90202 | Osc | At rest | P/P | B/B | |||
SpaceLabs 90207 | Osc | At rest | P/P | B/B | |||
In children | P/F | C/D | |||||
In elderly people (different posture) | P/P | A/C | |||||
In pregnancy | P/P | A/C | |||||
P/P | B/B | ||||||
P/P | B/C | ||||||
In pre-eclampsia | F/F | D/D | |||||
P/P | C/C | ||||||
During haemodialysis | P/P | C/B | |||||
SpaceLabs 90217 | Osc | At rest | P/P | A/A | |||
SpaceLabs 90207 | Osc | At rest | P/P | B/B | |||
In children | P/F | C/D | |||||
In elderly people (different posture) | P/P | A/C | |||||
In pregnancy | P/P | A/C | |||||
P/P | B/B | ||||||
P/P | B/C | ||||||
In pre-eclampsia | F/F | D/D | |||||
P/P | C/C | ||||||
During haemodialysis | P/P | C/B | |||||
SpaceLabs 90217 | Osc | At rest | P/P | A/A | |||
TM-2420/TM-2020 | Osc | At rest | F/F | D/D | |||
TM-2420 Model 5 | Osc | At rest | P/P | C/C | |||
TM-2420 Model 6 | Osc | At rest | P/P | B/B | |||
TM-2420 Model 7 | Osc | At rest | P/P | B/B | |||
TM 2421 - Takeda 2421 | Aus | In children (different posture) | A/B | ||||
Osc | C/C | ||||||
Takeda 2430 | Osc | At rest | P/P | A/A |