In vivo measurements of bone lead--a comparison of two x-ray fluorescence techniques used at three different bone sites

Phys Med Biol. 1989 Dec;34(12):1833-45. doi: 10.1088/0031-9155/34/12/007.

Abstract

In vivo bone lead measurements have been made on a group of about 120 people, most of whom were lead exposed workers. Two different x-ray fluorescence (XRF) techniques were used to make measurements at three bone sites. Finger lead was measured using 57Co sources, and lead measurements were made in both tibia and calcaneus with a technique based on 109Cd sources. The results of the bone lead measurements correlated strongly with each other and with the index of cumulative exposure, thus confirming the value and reliability of these in vivo measurements as a tool in the study of chronic lead exposure. Measurement precision, +/- 1 standard deviation, was highest for tibia +/- 7.4 micrograms (g bone mineral)-1, +/- 16.6 micrograms (g bone mineral)-1 for the calcaneus and lowest for phalangeal lead +/- 25.0 micrograms (g bone mineral)-1. Maximum absorbed doses to the skin were comparable for all three measurements (1-3 mGy). The mean whole body dose equivalents were all low, but that for the finger measurement, 0.1 microSv, was significantly less than for the calcaneus and tibia measurements 3-5 microSv.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Bone and Bones / analysis*
  • Calcaneus / analysis
  • Environmental Exposure
  • Fingers
  • Humans
  • Lead / analysis*
  • Male
  • Spectrometry, X-Ray Emission / methods*
  • Tibia / analysis

Substances

  • Lead