Improvements in the calibration of 109Cd K x-ray fluorescence systems for measuring bone lead in vivo

Phys Med Biol. 1994 Dec;39(12):2263-71. doi: 10.1088/0031-9155/39/12/009.

Abstract

A 109Cd K x-ray fluorescence (XRF) system using a point source in a back-scatter geometry is described. The suitability of plaster-of-Paris phantoms as targets for intercalibration standards was evaluated. When the phantom concentrations were measured by inductively-coupled-plasma mass spectrometry (ICPMs), the calculated phantom concentrations underestimated true concentrations by an average of 15%. Since calculated values are used to calibrate the K XRF system, in vivo bone-lead concentrations may be similarly underestimated. The difference between calculated and measured concentration is attributable to impurities in the plaster of Paris (e.g. calcium carbonate). The ICPMS-measured concentrations were used to calibrate the K XRF system. The same phantoms were also measured as 'unknowns' by a bone-lead measurement system (Abiomed, Danvers, MA, USA). The commercial system overestimated the lowest-concentration phantoms and underestimated the phantoms with concentrations above 15 microg Pb/g plaster of Paris. The commercial system and our system were compared by measurement of the new phantoms in air and in water. The K XRF system exhibited better precision in both situations. On the basis of this work, we recommend that plaster-of-Paris phantoms used to calibrate K XRF measurement systems be analysed first by ICPMS or another valid analytical technique.

Publication types

  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Bone and Bones / metabolism*
  • Cadmium Radioisotopes / analysis*
  • Calibration
  • Lead / analysis*
  • Lead Poisoning
  • Mass Spectrometry
  • Phantoms, Imaging
  • Spectrometry, X-Ray Emission / instrumentation*
  • Spectrometry, X-Ray Emission / methods*
  • Time Factors
  • X-Rays

Substances

  • Cadmium Radioisotopes
  • Lead