rss
Br J Ophthalmol 1978;62:145-154 doi:10.1136/bjo.62.3.145
  • Research Article

Effects of ultrasound on the corneal endothelium: II. The endothelial repair process.

Abstract

The responses of corneal endothelium to a 2-minute exposure of ultrasound from a clinical phacoemulsification device have been studied by specular microscopy and both scanning and transmission electron microscope. Healing of lesions occurred in two ways, and these were related to lesion size. Small lesions healed by a process of rosette formation, and large lesions were repaired by a process of endothelial dedifferentiation and migration. All damaged areas were re-covered with endothelial cells within 24 hours of exposure, and at this time corneal thickness had returned to control levels. The implications are discussed in relation to the phacoemulsification procedure.

This Article

Services

  1. Request permissions

Responses

  1. Submit a response
  2. No responses published

Social bookmarking

Register for free content


Free sample
This recent issue is free to all users to allow everyone the opportunity to see the full scope and typical content of BJO.
View free sample issue >>

Free archive
The full back archive is now available for BJO. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006, back to volume 1 issue 1.